National Instruments Corporation
37
NI 5402/5406 Calibration Procedure
11. Determine the error for each measurement using the following
formula:
DC Error
=
(
Measured DC Output Value
) – (
Ideal DC Output Value
)
12. Compare the DC error to the
calibration test limits
and
published
specifications
.
13. Abort waveform generation by calling the niFgen Abort
Generation VI.
14. Close the instrument driver session, destroy the instrument driver
session and all of its properties, and release any memory resources
NI-FGEN uses by calling the niFgen Close VI.
15. If any of the errors are greater than the
calibration test limits
, perform
a self-calibration procedure, followed by a reverification procedure.
LabVIEW Block Diagram
C/C++ Function Call
Call
niFgen_Abort
Generation
using the following
parameter:
vi
: The session handle
returned from
niFgen_init
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niFgen_close
using the following
parameter:
vi
: The session handle
returned from
niFgen_init
.