National Instruments Corporation
13
NI 5402/5406 Calibration Procedure
8.
Close the instrument driver session, destroy the instrument driver
session and all of its properties, and release any memory resources
NI-FGEN uses by calling the niFgen Close VI.
9.
If any of the errors are greater than the
calibration test limits
, perform
an external
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niFgen_close
using the following
parameter:
vi
: The session handle
returned from
niFgen_init
.