NI 5402/5406 Calibration Procedure
46
ni.com
12. Set the main DAC value by calling the niFgen Write Binary 16 Analog
Static Value VI.
13. Set the calibration ADC input by calling the niFgen Property Node and
selecting
Calibration»Cal ADC Input
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niFgen_Write
Binary16Analog
StaticValue
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
value
:
27,232
LabVIEW Block Diagram
C/C++ Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
""
(empty string)
attributeID
:
NIFGEN_
ATTR_CAL_ADC_INPUT
value
:
NIFGEN_VAL_
ANALOG_OUTPUT