National Instruments Corporation
49
NI 5402/5406 Calibration Procedure
21. Set the main DAC value by calling the niFgen Write Binary 16 Analog
Static Value VI.
22. Disable the analog output by calling the niFgen Property Node and
selecting
Basic Operation»Output Enabled
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Write
Binary16Analog
StaticValue
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
value
:
10,232
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViBoolean
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_OUTPUT_ENABLED
value
:
VI_FALSE