NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
19
WiFi
The NanoPhotometer
®
is equipped with WiFi, which can be used as a WiFi network or as a WiFi
Hotspot. The WiFi network allows data transfer and direct printing via AirPrint
®
/ IPP compatible
printers supporting PDF format.
Note
: IPP version 2.2 is required and some printer configuration settings might need to be changed in
order to allow communication with the NanoPhotometer
®
The WiFi Hotspot provides the option to control the NanoPhotometer
®
by other WiFi devices like
Windows computer, smartphones or tablets.
WiFi Hotspot connection details:
SSID
: NanoPhotometer
®
serial number
Password
: Implenuser
Note
: Due to the limitations of some handheld devices saving to a wireless device is limited to 40
measurements per dataset. Larger datasets can be saved on the NanoPhotometer
®
itself.
HDMI
There is an HDMI port located on the rear panel of the NanoPhotometer
®
which is compatible with
HDMI 1.4 cables (or better) to connect the NanoPhotometer
®
to HDMI compatible monitors.
Note:
The maximum HDMI cable length is 5 meters.
N
ANO
P
HOTOMETER
®
S
PECIFICATIONS
NanoVolume Performance
Detection Range dsDNA
1 ng/µl to 16,500 ng/µl (N50: 5 ng/µl to 7,500 ng/µl)
Detection Range BSA
0.03 mg/ml to 478 mg/ml (N50: 0.15 mg/ml to 217 mg/ml)
Minimum Sample Size
0.3 µl
Photometric Range
(10 mm equivalent)
0.02 - 330 A (N50: 0.1 – 150A)
Path length
0.67 and 0.07 mm
Dilution Factor
15 and 140
Vortexer
2,800 rpm; tube size up to 2.0 ml
Cuvette Performance
Detection Range dsDNA
0.1 ng / µl to 130 ng / µl
Detection Range BSA
0.003 mg / ml to 3.7 mg / ml
Photometric Range
0 - 2.6 A
Center Height (Z-Height)
8.5 mm
Cell Types
outside dimension 12.5 x 12.5 mm
Heating
37°C ± 0.5°C