NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
65
10. Select Replicates none, 2 or 3
11. Measure a blank and depending on the replicate selection, all required
concentrations. Absorbances of the replicates will be shown in the results
area and if replicates are selected a mean value for each standard. It is
possible to exclude single measurements from the curve calculation by
switching the toggle switch off.
Note
: After the first sample measurement the standard curve can no
longer be altered.
12. Once a Standard curve is created or loaded it will be used for
concentration calculations of measured samples. It may be necessary to
do a blank measurement.
13. Apply sample and press the sample button to initiate the sample
measurement.
Note
: Once the sample measurement is initiated it is no longer possible
to change the standard curve.