NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
81
6. Baseline correction is set to off as default. Enabling the baseline corrections shows a list
with different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823 nm
(N/A N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm)
7. Option to smooth the graph with different boxcars. Options: Off, 1 = boxcar 11 (default), 2 =
boxcar 21 and 3 = boxcar 61
8. Option to set/calculate a dilution factor for manual diluted samples.
9. Apply the blank ddH20 or buffer to the illuminated sample window for the
reference measurement and select blank to initiate the reading.
Note
: The illumination of the sample window can be switched off in the
preferences.
10. Use a lint-free laboratory wipe to clean both the sample window and
mirror prior to applying the next sample.
Note
: It can be helpful to apply the blank a second time and read it as a
sample to ensure a proper blank.