NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
70
Figure 1 Bacterial growth curve
M
EASUREMENT
P
ROTOCOL
1. Select the OD600 icon on the home screen
2.
To change between NanoVolume and cuvette application, use the
Change to Cuvette/NanoVolume button below the parameter area
(NP80/C40 only).
3. For
Cuvette Application
:
Select the path length depending on the used cuvette.
Options are: 0.5 mm, 1 mm, 2 mm, 5 mm and 10 mm
For
NanoVolume Application:
Select the dilution depending on the sample concentration
Note
: There is no automatic path length setting in this method. Select either a virtual dilution
of 15 (path length 0.67 mm) or of 140 (path length 0.07 mm)