7. Option to smooth the graph with different boxcars. Options: Off, 1 = boxcar 11 (default), 2
= boxcar 21 and 3 = boxcar 61
8. Option to set/calculate a dilution factor for manual diluted samples.
9. Apply the blank ddH
2
0 or buffer to the illuminated sample window for
the reference measurement and select blank to initiate the reading.
Note:
The illumination of the sample window can be switched off in
preferences.
10. Use a lint-free laboratory wipe to clean both the sample window and
mirror prior to applying the next sample.
Note:
It can be helpful to apply the blank a second time and read it as a
sample to ensure a proper blank.