NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
74
For NanoVolume application with submicroliter cell (C40 only):
Select Lid depending on your sample concentration (for concentration ranges of different
lids see page 39 Submicroliter Cell Measurements Basics).
Required sample volume for each lid is shown.
Note:
A new Blank is recommended when changing the dilution lid.
4. If it is desired to heat the sample to 37°C use the toggle switch to turn
on cell holder heater. When the cuvette holder has reached 37°C the
toggle color changes to green.
Note
: Only available for cuvette applications (NP80 and C40).
5. Enter desired wavelength (
λ
) to be measured. It is possible to measure up to 20
wavelengths simultaneously. More wavelength (
λ
) options can be added by selecting the
Add Wavelength button. Added wavelength can be deleted with
6. Baseline correction is set to off as default. Enabling the baseline corrections shows a list
with different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823 nm
(N/A N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm).