NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
63
For NanoVolume Application:
Select the dilution depending on the sample concentration
Note
: There is no automatic path length setting available in this method. Select either a
virtual dilution of 15 (path length 0.67 mm) or of 140 (path length 0.07 mm) depending on
your sample concentration.
For NanoVolume application with submicroliter cell (C40 only):
Select Lid depending on your sample concentration (for concentration ranges of different
lids see page 39 Submicroliter Cell Measurements Basics).
Required sample volume for each lid is shown.
Note:
A new Blank is recommended when changing the dilution lid.
5. If it is desired to heat the sample to 37°C use the toggle switch to turn on
cell holder heater. When the cuvette holder has reached 37°C the toggle
color changes to green.
Note
: Only available for cuvette applications (NP80 and C40).
6. The baseline correction is disabled as default. It is possible to choose between different
wavelengths for the baseline correction.
Options are: 377 nm, 604 nm, 650 nm, 770 nm (not N50) and 823 nm (not N50)
Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm)