NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 92 -
8.6.1.2.
Non-AXiiiS Workflow
The trajectory creation step in sec 8.5.4 should capture the final trajectory adjustment in this
case since it was aligned to the probe signal void. No further adjustment should be needed but
ensure that the software rendered LDP aligns to the LDP signal void when applicable.
CAUTION:
Final insertion to PDP should always be done by hand, under full user control
where tactile feedback is important.
•
Ensure the Interface Platform is properly attached to the AtamA system
•
Ensure the IP Power and Motor Plugs are inserted into proper receptacles
on the Interface Platform as shown in Figure 8.33 on page 87
Initiate Self-Test in the M*Vision software by selecting the
Start
button under
the
System Self-Test
heading on the right menu bar (See Figure 8.31). This
mode directs the user to hook up the Advanced Probe Driver (or RPD) / Laser
Delivery Probe and test functionality and then deliver the probe to the
established pre-insertion depth in the brain.
•
Depress the Laser Foot Pedal in the MRI control room
for two seconds prior to moving to Self-Test
Note
: Interface Platform Display will display the initiation
of Self-Test (see image right) prior to depressing (activating)
Foot Pedal. Display will automatically advance to the next
step after activation of foot pedal.
•
The balance of the Self-Test is completed within the MRI
scan room at the Interface Platform - the display will
guide the user through the required steps
•
If manual probe movement was selected at the start of the workflow step, the Self-Test
mode will skip any functions of the APD or RPD.
Note
: Self-Test may be skipped upon subsequent trajectories by
selecting the
Skip Additional Self-Test
icon (see image right). It is highly
recommended to perform a Self-Test any time a new Laser Delivery Probe is introduced during
a given procedure. A dialog box (below) will display the following upon selecting
Skip Additional
Self-Test.
Self-test Initiated…
Press the Foot Pedal for
2 Seconds
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...