NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 150 -
Once all desired treatment slices and angular positions of the probe have been used to optimize
the desired coagulation result, the user can select end treatment at any time. Any running MRI
acquisition should be stopped after all tissue returns to baseline temperature. Tissue further
away from the probe may take longer to return to near baseline temperature. Use the Pixel
Temperature Query tool and the temperature map to monitor temperatures of all tissue.
The probe can then be detached from the APD/RDP Follower by depressing the green locking
button and the patient can be prepared for burr hole closure by removing the probe from the
patient.
CAUTION:
The probe must first be detached from the APD/RPD Follower before
removing from the patient. Do not remove the probe from the patient while still
attached to the APD/RPD Follower.
Post treatment MR imaging such as blood sensitive GRE imaging may be used to identify any
immediate effect.
9.
Electrical Rack Requirements and Technical Data
•
Class 1 , with protective earth, protection against electric shock
•
Type CF degree of protection against electric shock
•
Input voltage: 120 V AC, 10 A, single phase, 50/60 Hz through a medical grade power
cord.
•
IPX0 degree of protection against ingress of water
•
Laser class 4
•
Inlet
CO
2
pressure gauges for each tank read >4500 kPa (>650 psi) to max 6000 kPa (880
psi)
•
The Electronics Rack provides continuous power to all internal rack components, to the
external Workstation (24 volts DC, medical grade) and to the Interface Platform (48 volts
DC and 12 volts DC, medical grade).
Table 9.1:
Electronics Rack Data
Electrical Connection Data
Voltage
120 VAC
Current
10 A
Number of Phases
1
Frequency
50/60 Hz
Class
Class I Equipment
International Protection rating
IPX0
UPS time out
After 8 minutes operating on battery power
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...