NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 142 -
•
Select the desired laser output by clicking on High, Medium or Low in the
Laser menu.
These settings represent different combinations of laser power and pulse
settings to achieve varying energy output levels.
WARNING:
To prevent patient injury, the user must confirm that the desired Laser Output
setting is selected before delivering laser energy.
8.8.3.3.
Initiate Delivery of Laser Energy to Tissue
•
Select
Mode
under the
Laser
heading to switch laser from
Standby
to
Ready
in preparation for energy delivery
The foot pedal is active only when laser is in
Ready
mode. Once the LDP has
cooled to its target range, the laser interlock switch will be released and the
status switches from
Standby to Ready
.
•
Depress foot pedal to initiate laser emission
When the laser foot pedal is depressed, displayed mode will be set to
Laser
to demonstrate a visual indicator of laser emission
and audible beeping will
be heard to
indicate that laser emission is active. If the foot pedal is
released, the laser mode will revert to
Ready
. The values under
Properties
will update current values when laser is operational.
CAUTION
: Do not initiate laser emission if the LDP is not inserted in tissue or before
the LDP connections are made.
CAUTION
: Ensure the desired LDP trajectory does not interfere with the MR bore or
NeuroBlate equipment prior to the insertion of LDP into tissue.
CAUTION:
Immediately release the laser footswitch upon the display of any error
condition.
•
Continue laser emission while monitoring thermal dose contours overlaid onto thermal
monitoring view-panes on the M*Vision display screen
M*Vision monitors operating conditions including LDP cooling to define when laser can
enter
Standby
mode. When operating conditions are normal and LDP has reached set-point cooling
level, system will release the laser interlock allowing
Standby
mode activation. The user must
enable energy delivery by switching to
Ready
mode via
Mode
button. Laser energy delivery is
controlled by user depression of foot pedal.
WARNING:
To prevent patient injury, do not mechanically clamp down the laser
footswitch during laser energy delivery.
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...