NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 16 -
1.2.5.
Advanced Probe Driver (APD) and Robotic Probe Driver (RPD)
Both the (single use) Advanced Probe Driver (APD) and (single use) Robotic Probe Driver (RPD)
are mounted to the NeuroBlate System Interface Platform (IP). The APD follower is designed to
be mounted onto a skull-mounted stereotactic anchoring device. The RPD follower is a lower
profile version of the APD follower designed to adapt to an appropriate skull bone anchor.
Other than the follower assembly, the functionality of APD and RPD is the same. The APD and
RPD Position Feedback Plug (Figure 1.8) connects to the Interface Platform to communicate the
Laser Delivery Probe’s position in the brain to the system user interface. The APD/RPD rotates
or translates (extends or retracts) the NeuroBlate Laser Delivery Probe. The APD/RPD provides
linear translation of up to 40 mm and total rotation of 340°.
The APD and RPD is comprised of two components: the Commander and the Follower (Figure
1.8), connected by an umbilical cable. The RPD uses a shortened version of the Follower as
shown in Figure 1.9. A Rotary Test Adapter is included with the APD and RPD, to be used
during Self-Test to simulate the attachment of Laser Delivery Probe during the procedure.
Figure 1.8:
Advanced Probe Driver Components
Rotary
Test Adapter
3 Thermal
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...