7.4.1.
Software Boot Up and Launch Screen
The LCD monitor in the Control Workstation will display the boot up sequence after system
power-on. When completed, the login display will appear (Figure 7.4.1 left).
•
Click the icon labeled
Site User
(Figure 7.4.1 left)
•
Type in the password provided by Monteris Medical
•
Confirm workstation PC time and date displayed on lower right side of the display matches
the current time and date of the MRI host computer to ensure the timescale of patient data
from the MRI is more accurately aligned to events within the NeuroBlate system
- Contact Monteris Technical Support if changes to the time setting are required
•
Double-click the
M*Vision for NeuroBlate
icon to launch the software application (Figure
7.4.1 right)
Figure 7.4.1: Boot Up Display (left) and Desktop Display (right)
NOTE:
The M*Vision™ software must be running before any DICOM data can be
sent to the Control Workstation from the MRI.
Upon start up, M*Vision launches in the
Start
task (see Figure 7.51 next page). The default task
,
requiring user selection of the MRI system being used before proceeding. The user can move
between the workflow tasks as noted by the icons in the upper right corner of the screen.
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...