![Monteris Medical Neuroblate Instructions For Use Manual Download Page 108](http://html1.mh-extra.com/html/monteris-medical/neuroblate/neuroblate_instructions-for-use-manual_1830188108.webp)
NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 108 -
Using the following steps, the sterile LDP can now be advanced into the APD/RPD Follower and
into the brain until the LDP Locking Interface comes in full contact with the Mating Adapter on
the Follower (see Figure 8.6.31).
CAUTION:
Final insertion to PDP should always be done by hand, under full user
control where tactile feedback is important.
Figure 8.6.31: Laser Delivery Probe as Inserted into the APD/RPD Follower
Figure 8.6.32: Left: Correct LDP Insertion Technique; Right:
Incorrect
LDP Insertion Technique
•
Slowly insert the LDP through the Probe Driver Follower and visually align the LDP thumb
grip to the alignment mark (grey line) on the Mating Adapter (Figure 8.6.31)
•
Continue insertion until fully engaged to the Follower Mating Adapter and locked in
position; an audible click will be heard
WARNING:
To prevent inadvertent unlocking or slipping of the LDP Depth Stop
during insertion, hold the LDP at the bottom-most portion of the Locking Interface
while inserting into the brain as shown in Figure 8.6.32 to prevent potential injury.
Correct finger
grip location
Mating Adapter
Alignment Mark
LDP Thumb Grip
LDP Fully
Engaged to
Follower
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...