NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 115 -
- This toggles the SendBuffered from ON to OFF
•
The
Switches Menu
shown in for items 8. - 12. shown in Figure 8.44 should now read:
8.
SendIMA
ON
9.
SendRAW
OFF
10.
SendRHP
OFF
11.
SendMeasData
OFF
12.
SendBuffered
OFF
CAUTION:
Do not modify any other settings within menus, including
5 (Switches)
menu.
•
Type: q (to exit Switches Menu)
•
Type: q (to exit ideacmdtool)
CAUTION:
For Siemens MRI software version 19, the values for the
Switches Menu
should
be reset to their original values at the end of the NeuroBlate procedure following the same
procedure outlined on pages 105 – 106 to prevent suspending the MRI software requiring
an MRI system reboot to clear.
Special Instruction for Siemens Aera/Skyra running software VD13; VE11 or higher
Before real time data transfer begins, the Siemens MRI must be disconnected from the PACS
network and directly connected to the NeuroBlate Main PC forming a direct peer-to-peer
network connection. In this case, both the NeuroBlate system and the Siemens MRI will be
disconnected from the hospital PACS network temporarily during treatment monitoring. The
MRI connection to the PACS network can be restored following treatment monitoring once real
time data transfer is no longer necessary. This step can occur before the case begins or just
before treatment monitoring since it will not disrupt normal DICOM data transfer between the
MRI and NeuroBlate.
For Siemens or IMRIS MRI, proceed to step
8.8.2 Position Thermal Monitoring Planes in
M*Vision
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...