Figure 8.7.1: Trajectory Confirmation
If Manual Probe Movement was selected at the M*Vision
Start
task
(manual mode), an additional
Probe Confirmation
header will be
visible in the right side menu. After either skipping or completion of
Self Test, the user must confirm that the Linear and Rotary readings
displayed in the probe status bar (lower right corner of software GUI)
match the physical position and orientation of the Laser Delivery
Probe in the brain. If a match is confirmed, the user must select the
Confirm Linear
and
Confirm Rotary
icons before selecting yes under
Confirm Trajectory
.
•
Adjust the software rendered LDP image by manipulating the
white grab handles to match to the LDP artifact on the acquired
image (see Figure 8.7.1 top).
•
When the software rendered LDP image matches the LDP artifact
on the screen, click
Yes
under the heading “Confirm Trajectory”
on the right M*Vision menu bar.
Note that once the trajectory is confirmed, the tip of the software
rendered probe capsule will turn to blue enabling the user to
robotically manipulate the linear position of the LDP (see Figure 8.7.1 bottom).
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...