NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 114 -
8.8.1.
ENABLE THE MRI FOR REAL TIME TRANSFER
Siemens or IMRIS MRI
Prior to thermal delivery and monitoring, the data transfer interface for any Siemens MRI must
be enabled following patient registration and immediately preceding thermal monitoring.
Note:
The MRI system will revert to the default configuration at the end of the NeuroBlate procedure.
CAUTION:
Do not enable the real time data transfer interface prior to this point in the
workflow. To prevent unintended application exit of the host MRI system control software.
•
In Advanced User Mode, Launch Command Tool Shell and Enable Data Transfer
Type:
Ctrl/Esc
Select:
Run
Type:
cmd
and click
OK
Type:
ideacmdtool<enter>
Type:
5
(
Switches…
)
•
Type in the site specific password if requested. The “Switches” Menu should appear in its
default mode as shown in Figure 8.44:
Figure 8.44: IDEA Command Tool Switches Menu (on MR host PC)
•
At the
Switches>
prompt, Type:
8
and then
Enter
- This toggles the
SendIMA
Switches from OFF to ON
•
At the
Switches>
prompt, type:
12
and then
Enter
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...