NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 73 -
Figure 8.19c: Example of RPD Depth Stop Calculation with a Skull Bone Anchor
The following table helps identify the target depth range for the probe sizes available when
used with a skull bone anchor with/without Robotic Probe Driver. The depths shown assume
the bone anchor is fully threaded into the skull and the top of the bone anchor (where IGS
derived depth is measured from) is ~30mm from skull’s outer surface.
Probe
Size
Depth
Stop, Min
Depth
Stop, Max
MIN Depth to
PDP using an
bone anchor
only from outer
skull
MAX Depth to
PDP using an bone
anchor only from
outer skull
MIN depth
with an bone
RPD
from outer
skull
MAX depth
with an bone
RPD
from outer
skull
000
51mm
71mm
103mm
123mm
8mm
68mm
00
72mm
92mm
124mm
144mm
29mm
89mm
1
93mm
113mm
145mm
165mm
50mm
110mm
2
114mm
134mm
166mm
186mm
71mm
131mm
3
135mm
155mm
187mm
207mm
92mm
152mm
4
156mm
176mm
208mm
228mm
113mm
173mm
5
177mm
197mm
229mm
249mm
134mm
194mm
If using trajectory guidance/anchoring devices other than those available from Monteris
Medical please refer to the respective manufacturer’s Instructions for Use.
L (mm)
Probe Depth Stop Setting = L - 27mm
PDP x
OR Preparation Step illustration with RPD and
a standard bone anchor
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...