NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 76 -
Attachment Method 1– AtamA Attachment Interface
1.
Ensure AtamA patient board is fully installed and secure on the MRI table.
2.
Slide the arm ends (furthest away from the interface platform) into the IP Arm interface
completely until the “stop rings” touch the IP arm interface. (
Note:
A thumb screw is
provided on each side to lock the arms in place. Do not over tighten these screws as
they may damage the arm.)
Figure 8.41: Interface Platform (IP) to AtamA Connection
WARNING:
Do NOT lower the MRI table after the IP is attached to the AtamA Board. Doing
so may lead to patient injury or equipment damage.
IP
Power
Button
APD / RPD
Attachment
Location
IP
Arm
IP Arm
Interface
IP
Arm
AtamA
Board
IP Arm
Interface
“Stop Rings”