CAUTION:
Never pull on the thinner tubular or cable sections of the laser, cooling,
temperature sensor, and probe detection lines, as this may damage the components.
CAUTION:
The Laser Delivery Probe is fragile. Handle with care.
Figure 8.6.32: Adjustment and Setting of Depth Stop
•
Ensure LDP is fully inserted into the provided ruler/protective cover as shown in Figure
8.6.32 (bottom) - an audible click should be heard if reinserting the cover into the LDP
Locking Interface
•
Squeeze the tapered part of both yellow buttons to unlock the depth stop.
•
Carefully slide the Locking Interface so that the LDP tip aligns with the required distance
measurement on the ruler
•
Squeeze both (yellow) locking buttons at their wider end (see Figure 8.6.32); an audible click
should be heard when locked
•
Confirm depth stop is locked by gently pushing and pulling on the upper portion of the
locking interface - avoid manipulating the stops unless a change to probe depth is required
•
Recheck to ensure that the proper depth is set by matching the LDP tip to ruler graduations
and that the depth stop is locked prior to inserting the LDP into the brain
Squeeze (yellow)
buttons at wider end
to lock Depth Stop
Slide Locking Interface to adjust depth
130
13
5
14
0
145
Probe Tip
Depress (green) release button to remove the Ruler
Squeeze (yellow)
buttons at tapered
end to release
depth stop
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...