NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 90 -
8.6.
Self-Test, Probe Driver Attachment and LDP Insertion
8.6.1.
Final Trajectory Adjustment in
Insert
8.6.1.1.
AXiiiS Workflow
The selected trajectory should align to the MR visible wand if the AXiiiS is properly aligned to
the intended trajectory (Figure 8.6.11). If there is misalignment within this task, the selected
trajectory should be adjusted to align to the MR wand using the white grab handles as was
done in the
Trajectories
workflow step. This will ensure that the software rendered LDP will
align to the intended path during LDP insertion. During final adjustment of the trajectory, the
Point of Deepest Penetration (PDP) location must be set to define the deepest allowable point
for delivery into the brain.
CAUTION:
PDP LOCATION -This position defines the depth setting of the LDP and thus the
deepest point of penetration of the probe tip during probe insertion. During MR guided
probe insertion, the probe can begin treatment from any depth setting along trajectory. The
PDP must be defined at a safe depth location since probe linear motion can extend to this
point.
Figure 8.6.11:
Insert Task using the SFP and
Self-Test Initiation
PDP
Wand
PDP
Summary of Contents for Neuroblate
Page 1: ...Instructions for Use ...
Page 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Page 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Page 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Page 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...