92
Is the voltage
output load resistance at
least 2 kΩ and the current output
load resistance not more
than 510 Ω?
Rotation speed analog output
check
Are the connections
correctly configured?
Check the rotation speed analog
output settings.
Are they set correctly?
Yes
Yes
Yes
Contact us and describe the issue and
situation in as much detail as possible.
No
No
No
Connect correctly in accordance
with [2-6. Rotation Speed Analog
Output Connection].
Use a load resistance to suit the
specifications.
Set correctly in accordance with
[10. Analog Output].