25
Selecting Span Point Fine-adjustment Mode
Span point fine-adjustment mode can be selected from measurement mode as follows:
appears in the upper part when the
key is
depressed for approximately 2 seconds in measurement mode.
The upper display reads
when the
key is
pressed three times.
The upper display reads
and the reading on the lower
display flashes, indicating span point fine-adjustment mode has been
selected when the
key is pressed.
The lower display displays the torque value currently entered.
Apply an actual torque to the torque transducer, and adjust the display
to the actual torque value.
: Increments the torque.
Hold down the key to increment the torque continuously.
: Decrements the torque.
Hold down the key to decrement the torque continuously.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next
step.
Press the
key after setting.
is displayed after span point fine-adjustment.
Press the
key to exit span point fine-adjustment mode.
The system switches to measurement mode and the data set is
registered.
The change polarity setting will be temporarily canceled for zero/span fine-adjustment. (Restored
after adjustment)
The A/Z function is canceled once the zero/span fine-adjustment function ends.
Measurement
mode
Current torque
value
(3 times)
Measurement
Mode