46
Function List
Function No.
Item
Setting
value
Details
F-01
Decimal point display position
0
No decimal point
1
1000.0
2
100.00
3
10.000
4
1.0000
F-02
Torque display units
0
N
m
1
kN
m
2
V
F-04
Display frequency
0
4 cycles/s
1
20 cycles/s
F-05
Torque analog filter
0
1 Hz
1
10 Hz
2
30 Hz
3
50 Hz
4
100 Hz
5
300 Hz
6
500 Hz
7
1 kHz
8
6 kHz (Cannot be selected when F-54 is 0)
F-06
Torque digital filter
00 to 88
No. of moving-average samples = 2
m
+2
n
m: 10
1
setting, n: 10
0
setting
00
F-07
Rotation speed analog filter
0
1 Hz
1
10 Hz
F-08
Rotation speed digital filter
00 to 88
No. of moving-average samples = 2
m
+
2
n
m: 10
1
setting, n: 10
0
setting
00
F-09
Torque stabilization filter
No. of moving-average samples
0
No. of moving-average samples: 1
1
No. of moving-average samples: 2
2
No. of moving-average samples: 4
3
No. of moving-average samples: 8
4
No. of moving-average samples: 10
5
No. of moving-average samples: 12
6
No. of moving-average samples: 14
7
No. of moving-average samples: 16
8
No. of moving-average samples: 32