78
RS-232C/422/485 Communication Data Format
Command Mode Data Format
The address is fixed at [00] for the RS-232C interface.
Load data is entered as right-aligned data.
The sign of the data is indicated by [-] if negative and [+] if positive.
Load data is subject to zero suppression.
If an overload occurs, [OL] is output to the load data unit for torque and [OS] is output for rotation
speed.
All blank areas are spaces.
If decimal point is enabled, a decimal point is added at the specified location when decimal points
are specified on the calibration screen.
The unit part uses the value set by function number F-02. [0] →N·m, [1] →kN·m, [2] → V
The unit part for command 21 is fixed at [0], and the unit part for command 22 is fixed at [Space].
Data reading (Host → OPT-564)
For command numbers 20, 21, 22
2
4
6
@
0
0
2
0 Terminator
Return (OPT-564 → Host)
2
4
6
8
10
12
14
@
0
0
2
0
Unit
Terminator
2
4
6
8
10
12
14
@
0
0
2
0
-
1
0
Terminator
2
4
6
8
10
12
14
@
0
0
2
0
+
0
.
1
0
Terminator
2
4
6
8
10
12
14
@
0
0
2
0
+
O
L
0
Terminator
Command No.
Operation
20
Torque display read-out
21
Rotation speed display read-out
22
A/D data read-out
23
Frequency data read-out
Address
00 to 31
Address
00 to 31
Address
00 to 31
Address
00 to 31
Command No.
Address
00 to 31
Command No.
Data sign
Data