The display flashes for approximately 2 seconds when [-90 % of maximum
display value] is exceeded during symmetry correction.
The display flashes for approximately 2 seconds when the torque value input
does not exceed the previously set value for the positive setting or when the
torque value input does not exceed the negative side maximum display value for
the negative setting during linearization correction.
The display flashes for approximately 2 seconds when the torque value input is
not less than the maximum display value for the positive setting or when the
torque value input is not less than the previously set value for the negative setting
during linearization correction.
When F-54 = 0:
Illuminates when the input torque value exceeds 16 kHz in monitor mode.
When F-54 = 1:
Illuminates when the input torque value exceeds 384 kHz in monitor mode.
When F-54 = 0:
Illuminates when the input torque value is less than 4 kHz in monitor mode.
When F-54 = 1:
Illuminates when the input torque value is less than 96 kHz in monitor mode.
When F-54 = 0:
Illuminates when the input torque value exceeds [+110 % of maximum display
value] or exceeds 16 kHz in measurement mode.
When F-54 = 1:
Illuminates when the input torque value exceeds [+110 % of maximum display
value] or exceeds 384 kHz in measurement mode.
When F-54 = 0:
Illuminates when the input torque value is less than [-110 % of maximum display
value] or is less than 4 kHz in measurement mode.
When F-54 = 1:
Illuminates when the input torque value is less than [-110 % of maximum display
value] or is less than 96 kHz in measurement mode.
Illuminates when the rotation speed exceeds 27,500 in measurement mode.
Illuminates when the rotation speed is less than -27,500 in measurement mode.
Illuminates when no torque signal is input in measurement mode or monitor
mode.
Illuminates when a torque transducer error occurs. (With compatible TMHSB
torque transducer)
-