27
Linearization Correction Function
Torque value errors may arise between the zero and span (torque values) due to factors such as coupling
effects after calibration.
The digital linearization function minimizes measuring errors by correcting up to five points (five points
each for clockwise and counter-clockwise directions), excluding zero and span.
Selecting Linearization Correction Mode
Linearization correction mode can be selected from measurement mode as follows.
appears in the upper part when the
key is
depressed for approximately 2 seconds in measurement mode.
The upper display reads
when the
key is
pressed five times.
The upper display reads
and
flashes
on the lower display, indicating that linearization correction mode has
been selected when the
key is pressed.
Use the
or
key to set the required value on the lower
display.
0:
Counter-clockwise direction
1:
Clockwise direction
/
: Alter the value of the digit to be altered.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next
step.
Press the
key after setting.
D
isp
la
y
D
isp
la
y
Actual torque
value
Error
Span
0
Zero
Actual torque
value
0
Zero
Span
Correction point 3
Correction point 2
Correction point 1
After linearization correction
Before linearization correction
Example using
3 correction
points
(5 times)
Measurement
mode
To next page [A]