44
Calibration Data Selection
This equipment can save up to four sets of calibration data. (F-59)
Setting F-59 to 0 allows calibration data to be selected using an external control input.
Setting range: 0 to 4
F-59 SEL2 SEL1 Calibration
number
0
OFF
OFF
1
The calibration number is determined by the external control
input when F-59 = 0.
0
OFF
ON
2
The calibration number is determined by the external control
input when F-59 = 0.
0
ON
OFF
3
The calibration number is determined by the external control
input when F-59 = 0.
0
ON
ON
4
The calibration number is determined by the external control
input when F-59 = 0.
1
Ignore Ignore
1
The calibration number is determined by the number set when
F-59 = 1 to 4.
2
Ignore Ignore
2
The calibration number is determined by the number set when
F-59 = 1 to 4.
3
Ignore Ignore
3
The calibration number is determined by the number set when
F-59 = 1 to 4.
4
Ignore Ignore
4
The calibration number is determined by the number set when
F-59 = 1 to 4.
For example, if you have two TMHS torque transducers, the calibration data for the first transducer can
be saved to Calibration 1 and the data for the second transducer can be saved to Calibration 2,
simplifying the calibration procedure when calibrating the transducers.
By default, this is set to [0]. (Selected by external control input)
[Function F-21 maximum torque analog output display value], [Function F-22 maximum rotation
analog output display value], [Symmetry correction function], and [Linearization correction function]
are saved respectively for calibration data 1 to 4.
Calibration Prohibition
Calibration data set using the function F-59 can be locked to prevent accidental calibration. (F-97)
The default setting is for calibration of all data to be allowed. Note that calibration prohibition applies to
the following setting digits.
Setting range: 0000 to 1111
0: Allowed
1: Prohibited
10
0
digit: Calibration data 1
10
1
digit: Calibration data 2
10
2
digit: Calibration data 3
10
3
digit: Calibration data 4
Symmetry Correction Clear
Data corrected using the symmetry correction function can be cleared. (F-56)
Only the currently selected calibration data will be cleared.
Pressing the
key while
is displayed causes the lower display
indication to flash. To stop clearing the symmetry correction at this point, press
. Symmetry
correction clearing is not performed, and the system returns to measurement mode.
If the
key is pressed while
is flashing, the display changes to
,
and the symmetry correction clear operation ends.