28
The upper display reads
and the lower display
reads
.
Use the
or
key to alter the lower display to the
number of points to be corrected. (Up to five points can be set.)
0 to 5: Number of linearization correction points
The upper display reads
, and the lower display
shows a numerical value corresponding to the torque value
currently input. Pressing the
key with the actual torque to
be corrected applied to the torque transducer causes the lower
display to flash. Adjust the indicated value to the actual torque to
be corrected.
For example, if the theoretical actual torque to be corrected and
applied to the torque transducer is set to [1000], the target value
for adjustment here will be [1000].
: Increments the torque.
Hold down the key to increment the torque
continuously.
: Decrements the torque.
Hold down the key to decrement the torque
continuously.
: Cancels the setting and returns to measurement
mode.
: Registers the value displayed and proceeds to the
next step.
Press the
key after setting.
If there are two or more correction points, the upper display reads
. Continue with linearization correction using the
same procedure.
is displayed once linearization correction is
complete.
Press the
key to exit linearization correction mode. The
system switches to measurement mode and the data set is
registered.
[A]
If there are
two or more
correction
points,
perform the
operation
for the
number of
points.
If there is one
correction point
Measurement
mode