58
Analog Output Actual Torque Fine Adjustment Function (VADJ)
This function adjusts the analog output by applying the actual torque to the torque transducer.
Analog output actual torque fine-adjustment mode can be selected from measurement mode as follows:
appears in the upper part when the
key is
depressed for approximately 2 seconds in measurement mode.
The upper display changes to
when the
key is pressed eight times.
When the
key is pressed, the upper display reads
, and
flashes on the lower display.
The lower display changes to
when the
key is pressed.
Pressing the
key again causes the currently input torque
value to flash.
Set the torque transducer to the initial torque state, then adjust
while monitoring the analog output.
: Increments the analog output.
Hold down the key to increment the rotation speed
continuously.
: Decrements the analog output.
Hold down the key to decrement the rotation speed
continuously.
: Cancels the setting and returns to measurement
mode.
: Registers the value displayed and proceeds to the
next step.
Press the
key after setting.
The lower display changes to
when the
key is pressed.
Pressing
again causes the currently input torque value to
flash.
Adjust while monitoring the analog output.
After analog output
adjustment
After analog output
adjustment
Measurement
mode
(8 times)