4
Electromagnetic compatibility
4.1 Compliance details
As noted in section 3, the miha bodytec m.ove EMS training device
complies with IEC 60601-1-2:2014 and IEC 60601-2-10:2012.
According to these standards, the requirements apply as listed in
the tables below.
Guidance and manufacturer’s declaration – electromagnetic emissions
The miha bodytec m.ove EMS training device has been designed for use in electromagnetic environments
as outlined below.
The buyer or user of the miha bodytec m.ove device must ensure that the device is operated under such
conditions.
Emission tests
Compatibility
Electromagnetic environment –
guidance
RF emissions
CISPR 11
Group 1
The miha bodytec m.ove EMS
training device uses low RF power
for internal operation. Generating
very low RF interference emis-
sions, it is not likely to interfere
with electronic equipment in its
vicinity.
RF emissions CISPR 11
Class B
The miha bodytec m.ove device is
suitable for usage in any building,
including residential buildings and
those connected up direct to low-
voltage power supply networks
installed for residential buildings.
Harmonic distortion.
IEC 61000-3-2
Class A
Voltage fluctuations and flicker.
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
The miha bodytec m.ove EMS training device has been designed for use in electromagnetic environments
as outlined below.
The buyer or user of the miha bodytec m.ove device shall make sure that the device is used under such
conditions.
Immunity test
IEC-60601 test level
Compliance level
Electromagnetic envi-
ronment – instructions
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV on contact dis-
charge
±2 kV, ±4 kV, ±8 kV,
±15 kV air discharge
±8 kV contact discharge
±2 kV, ±4 kV, ±8 kV,
±15 kV air discharge
The miha bodytec m.ove
EMS training device can
be used in rooms without
humidity control. Floors
can be covered with syn-
thetic material. If so, the
relative humidity should
be at least 5 %.
Electromagnetic compatibility
2021-07-28
miha-bodytec m.ove EMS training device
42