Personnel:
n
Athlete/Patient
n
Trainer
The training session can be interrupted between individual exer-
cises to pause or discuss the next exercise with the trainer.
1.
Press the main controller (Fig. 57/14).
ð
The main level jumps to “0” and all LEDs are flashing.
The training time continues running during the pause.
2.
Turn the main controller to end the pause.
ð
The training session continues as usual.
Before starting with the workout (countdown) and during the pulse
phases, the electrodes connected via the channel outputs are
detected.
If a pair of electrodes is not detected before starting with the
workout (countdown), the corresponding channel is shown in gray
on the display and cannot be reactivated during the training ses-
sion. If this happens even though something was connected to this
channel, remove the transponder card again right away at the start
of the workout. This ends the training session. Next, check to make
sure that the contacts are properly connected and the electrodes
are sufficiently moistened. Correct the problem, if necessary.
If the contact of a channel is interrupted unexpectedly during the
workout, the intensity of the channel drops to “0” and cannot be
increased again afterwards. The channel is shown in gray on the
display. The channel intensity can only be increased again once
the electrode is properly reconnected.
Pausing the training session
Electrode detection/channel shutoff
Overview of programs/training plans
2021-07-28
miha-bodytec m.ove EMS training device
96