Replacing the Chamber Cap O-Ring
Fine fibers and particles between the O-ring and its sealing surfaces can
cause leaks, as can scratches or cuts in the O-ring or in the metal surfaces.
When it is necessary to replace the O-ring, follow these steps.
1.
Remove the O-ring from its groove by pushing a sharp tool into the ring
and prying it out.
Be careful not to scratch the metal surface of the chamber cap.
Scratches could result in an imperfect seal.
Figure 5-2. Removing the Chamber Cap O-Ring
AccuPyc 1330
Maintenance
September 1996
5-15
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...