Manually Controlling the Valves
This function enables you to perform a manual analysis by controlling the
valves.
Press
to enter manual mode.
Display
Entry
When Manual is shown in the display, you
can manually open and close the fill,
expansion, and vent valves by pressing the
appropriate keys.
7 - Opens and closes the fill valve
8 - Opens and closes the expansion valve
9 - Opens and closes the vent valve
Observe the indicators above the
,
, and
keys. When a valve is
open, the indicator is on. When a valve is
closed, the indicator is off.
Press
when you wish to exit manual
mode and return to display mode.
Prolonged opening of all three valves at one time may cause excessive
loss of helium.
1
Manual
p = (pressure)
FILL
EXPAND
VENT
SAVE
AccuPyc 1330
Manually Controlling the Valves
September 1996