Calibrating the Pycnometer
You should check or calibrate the pycnometer anytime you restart it. For a
quick check, run an analysis with an empty cup to see how close the average
volume is to 0. It should be
±
0.05% of full-scale. If it is not within
±
0.05%
of full-scale, calibrate according to the instructions given below. When you
calibrate the pycnometer, cell volume and expansion volume in the set-up pa-
rameters are updated automatically.
When calibrating the pycnometer, you should set up your calibration pa-
rameters so that 10 purges and 10 runs are performed. See the section
on Analysis or Calibration Parameters later in this chapter.
1.
Place an empty cup in the cell chamber.
2.
Replace the cell chamber cap.
3.
Press
to begin the calibration procedure.
Display
Entry
Enter the volume of the calibration standard
and press
.
Range:
0.1 to 999.0000 cm
3
Press
to begin the calibration or
to cancel the calibration.
Status messages are displayed during the first
phase of calibration. When the first phase
ends, the pycnometer beeps three times.
Place the calibration standard* in the cup in
the cell chamber.
In most cases, use one ball for the 100-cm
3
pycnometer; two balls for
the 10-cm
3
pycnometer. For best results, sample volume should be ap-
proximately equal to the volume of the calibration standard. Therefore,
if you are calibrating the 10-cm
3
pycnometer for use with smaller sam-
ple volumes, you may wish to use only one ball.
Replace the cell chamber cap; then press
. When you press
, calibration
continues and operational status messages are
continually displayed.
•
Volume of cal std:
1.0000 cm
3
ENTER
[Enter] to start
[Escape] to cancel
ENTER
CLEAR
Insert cal std
[Enter] to start
ENTER
ENTER
AccuPyc 1330
Calibrating the Pycnometer
September 1996
4-15
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...