Messages Displayed During an Automatic Operation
When an automatic operation is in progress, status messages are continually
displayed.
Automatic operation
has been canceled
Cause:
The automatic operation has been canceled by the user.
Action:
Wait for the termination process to complete or end the
termination process by pressing
two
more times.
Print in progress
[ESCAPE] to cancel
Cause:
Status message displayed when a report is being printed.
Action:
None; this is a status message only.
Printer port
not responding
Cause:
The printer took longer than 30 seconds to acknowledge
receipt of data from the pycnometer.
Action:
Check to make sure the printer is properly connected to
the pycnometer, is turned on, and is on line.
Sending line (line number)
[ESCAPE] to cancel
Cause:
Status message displayed when data are being transmitted.
Action:
None; this is a status message only.
Transmission port
not responding
Cause:
The receiving device took longer than five seconds to ac-
knowledge receipt of data from the pycnometer.
Action:
Make sure the receiving device is properly connected to
the pycnometer RS-232 port and is turned on. Verify that
the serial I/O parameters controlling the receiving device
correspond with the data transmission parameters in the
set up mode.
CLEAR
5-12
March 2001
Error and Status Messages
AccuPyc 1330
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...