10. Remove the screws securing the module mounting rails to the baseplate.
Figure A-5. Module Mounting Rails
11. Grasp the module and pull it straight up. A convenient method of removing
the module is to replace the cap completely and use the knob as a handle to
pick up the module. Observe the standoffs mounted to the baseplate and the
corresponding recesses in the rails.
12. Disconnect the cable assemblies going to the valves mounted to the block.
Figure A-6. Disconnecting Cable Assemblies
Mounting Rails
Standoffs
Disconnect Valve
Cables Here
Disconnect Electronics
Module Cables Here
APPENDIX A
AccuPyc 1330
A-4
September 1996
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...