Selecting Units of Pressure Measurement
and Operating Language
The units of measurement for pressure may be displayed in pounds-per-
square-inch gauge (psig) or kilopascal gauge (kPag). In this manual, psig is
used in the examples shown. The pycnometer may be operated in any of five
languages: English, French, German, Italian, or Spanish. The default for units
is psig and the default language is English. If you wish to use these defaults,
no further action is necessary; proceed to Greasing the Chamber Cap
O-Ring. If you wish to change the defaults:
1.
Press
, then press
until Unit types is displayed.
Press
.
2.
Press
then
to change the value to kPag.
3.
The Language prompt is displayed. Press
until the desired lan-
guage is displayed, then press
.
If there are error messages in the message queue (indicated by an aster-
isk next to the Reload prompt), delete the messages before changing lan-
guages. Refer to Error, Report, and System Status Messages, Chapter 4.
Make sure you press
after changing languages.
4.
Place the appropriate template (included in the accessory kit) over the
pycnometer keypad.
2
CHOICE
ENTER
CHOICE
ENTER
CHOICE
SAVE
SAVE
AccuPyc 1330
Equipment Setup
September 1996
2-9
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...