What Happened
Why
What To Do
Specified pressure not
reached or maintained.
(continued)
Helium tank empty.
Check tank. Minimum
recommended pressure
is 200 psig.
Cylinder Shut Off
Valve closed.
Check that valve is
open.
Pressure regulator
defect or pressure set
too low.
Check regulator to be
sure it is set properly.
Call a Micromeritics
Service Representative,
if necessary.
Zero offset (of pressure
transducer) too low.
Check offset by
opening chamber cap.
If offset is negative,
run new zero offset
(Zeroing the Pressure
Transducer, Chapter 4).
Helium drained from
tank.
Leaks in the gas line
connection.
Pressurize the system.
Close, then open the
Cylinder Shut-Off
Valve. If the needle on
the pressure gauge on
the gas cylinder jumps
abruptly, a leak in the
gas line connections
may be indicated.
Check all gas line
connections. (Refer to
Rear Panel
Connections in
Chapter 2 for
connection instructions.)
Pycnometer was left in
manual mode with all
the valves open or the
fill valve open and the
chamber cap off.
Close all valves, then
attach a new tank of
helium.
AccuPyc 1330
Troubleshooting
September 1996
5-3
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...