Transmission port
waiting for Xon
Cause:
The receiving device stopped transmission by sending an
Xoff, and hasn’t resumed the transmission by sending an
Xon.
Action:
None; when the receiving device is ready for more data,
it should send the pycnometer an Xon.
The following message is displayed on the printer or on the screen for trans-
mitted data.
No collected data to report, or all runs excluded
Cause:
This message is printed or transmitted if you requested a
report but there is either no data available or data has
been excluded via review mode.
Action:
Initiate an automatic operation or return to review mode
and include at least one run.
AccuPyc 1330
Error and Status Messages
September 1996
5-13