Replacing the Gas Inlet Filters
The gas inlet contains a filter that may become contaminated after extended
periods of use. If you are getting very low or erratic values for volume and
density, the filter may be contaminated. Replace it as follows.
1.
Turn off the gas supply at the regulator mounted to the cylinder by turn-
ing the cylinder shut-off valve fully counterclockwise.
Figure 5-4. Turning Off Gas Supply
2.
Press
on the pycnometer keypad to enter manual mode.
Then press VENT to open the vent valve, EXPAND to open the expan-
sion valve, and FILL to open the fill valve.
The power cord should be disconnected from the unit before disconnect-
ing the gas supply. Failure to do so could result in electrical shock.
3.
Observe the pressure reading. When the pressure drops to less than 1
psig, place the power switch on the back of the pycnometer in the off
(0) position. Remove power cord.
4.
Disconnect the gas supply line from the rear panel of the pycnometer by
turning the nut on the INLET connect counterclockwise with a 7/16-in.
open-end wrench.
5.
Disconnect the gas vent line (if attached) by gently pulling the flexible
tubing from the rigid tubing labeled VENT at the rear of the pycnometer.
1
Cylinder Shut-
Off Valve
Maintenance
AccuPyc 1330
5-18
September 1996
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...