MAN_ERR:
Pressure overrange
Cause:
The fill valve was left open until the maximum system
pressure was exceeded.
Action:
Close the fill valve and open the vent and expansion
valves. Allow the pressure to stabilize.
PRT_ERR: Timeout
failed to respond
Cause:
The printer took longer than 30 seconds to acknowledge
receipt of data from the pycnometer.
Action:
Check to make sure the printer is properly connected to
the pycnometer, is turned on, and is on line.
SYS_ERR: Power Fail
NN/ZZ Runs Completed
Cause:
A power failure occurred and when power resumed, the
automatic operation was canceled.
Action:
Restart the automatic operation if desired.
TRN_ERR: Timeout
failed to respond
Cause:
The receiving device took longer than 10 seconds to ac-
knowledge receipt of data from the pycnometer.
Action:
Make sure the receiving device is properly connected to
the pycnometer RS-232 port and is turned on. Verify that
the serial I/O parameters controlling the receiving device
correspond with the data transmission parameters in the
set up mode.
USR_ERR: Cal std
10% of full-scale
Cause:
You tried to enter a calibration standard volume that is
less than 10% of the nominal full-scale volume.
Action:
Enter a volume that represents at least 10% of the nomi-
nal cell chamber volume.
5-10
March 2001
Error and Status Messages
AccuPyc 1330
Summary of Contents for AccuPyc 1330
Page 1: ...AccuPyc 1330 Pycnometer Operator s Manual V3 03 Part No 133 42808 01 January 2005 ...
Page 6: ...CHAPTER 1 GENERAL INFORMATION Organization of the Manual Conventions Equipment Description ...
Page 83: ...CHAPTER 6 ORDERING INFORMATION ...
Page 86: ...APPENDIX A CHANGING THE ANALYSIS MODULE ...
Page 92: ...APPENDIX B ANALYSIS THEORY ...
Page 96: ...APPENDIX C CALIBRATION THEORY ...
Page 99: ...APPENDIX D FORMAT OF TRANSMITTED DATA ...
Page 108: ...APPENDIX E PRINTER CHARACTER SET ...
Page 111: ...INDEX ...