Page 15
Functional Principle, Technical Data
combiSENSOR 64x0
2.3
Technical Data
Controller
KSS6420
KSS6430
KSS6420(01)
KSS6430(01)
Sensor
KSH5(01)
KSH10
Target thickness insulator
40 µm ... 3 mm
40 µm ... 6 mm
Resolution
1)
static, 104 Sa/s
0.0018 %
0.0004 %
0.0030 %
0.0006 %
dynamic, 3.9 kSa/s
0.0075 %
0.0015 %
0.0120 %
0.0025 %
Bandwidth
analog: 1 kHz (3 dB)
2)
, digital: 2.6 … 3900 Sa/s (adjustable)
Linearity
± 0.05 %
Temperature stability
4)
Sensor
± 50 ppm
(+10…+60 °C)
Temperature stability
Controller
± 50 ppm
± 50 ppm
± 50 ppm
± 70 ppm
Temperature range
controller: +10…+50 °C; 10…+50 °C; sensor cable: -10…+125 °C
Power supply
12…36 VDC (5.5 W)
Output
Analogue
capacitive, eddy current and differential signal: 0-10 V (short circuit proof);
internal sensor temperature signal (not scaled)
Ethernet
capacitive, eddy current, differential and internal temperature signal: 24 Bit
EtherCAT
capacitive, eddy current, differential and internal temperature signal: float
Trigger
TTL, 5 V
Target geometry
Straight surface or min. diameter 200 mm
3)
Working distance
2 ... 5 mm
4 ... 10 mm
Min. diameter measurement surface
45 mm
65 mm
Protection class
sensor: IP 54, controller: IP 40
Weight
sensor: 80 g ; controller: 750 g
Storage temperature
sensor, cable: -10…+100 °C, controller: 0…75 °C
1)
Difference signal of the digital output, measured with working distance = 50 % FSO
2)
Only valid when sampling rate = 3900 Sa/s
3)
Reference material ground electrode: VA steel (1.4571) or aluminium.
Changes of the ground electrode (material or geometry) require a recalibration of sensor and controller by the manufacturer.
4)
Circumferential clamping starts 5 mm behind the sensor surface, see
Fig. 7
.