Page 33
Ethernet Interface
combiSENSOR 64x0
By default, the IP address of the controller is
set to 169.254.168.150. Communication with
the controller is done on the data port 10001 for
measurement transmission. A command port
(Telnet, port 23) is used for sensor commands.
The IP settings and the data port can be changed at any time:
- by using the web browser. Enter the current IP address into the address bar. Go to the menu
Settings
> Digital Interfaces > Ethernet settings
to set a new IP address, activate DHCP or
change the data port.
- by using software commands, see Chap.
6.4
.
- by using the
SensorFinder
software.
If you activate DHCP, access to the controller via a DHCP host name is possible. The host name contains the
device name and serial number. Structure: NAME_SN e.g. KSS6420_1001.
The controller supports UPnP. If you use an operational system with activated UPnP client e. g. standard with
Windows 7, the controller is listed in the explorer as a device automatically. This is helpful, if you do not know
the IP address of the controller.