Page 46
Ethernet Interface
combiSENSOR 64x0
Direct connection to PC, controller with static IP (Factory setting)
Network
PC with static IP
PC with DHCP
Controller with dynamic IP, PC with DHCP
Connect the controller to a PC using a direct Ethernet connection (LAN).
Use a LAN cable with RJ-45 connectors for this.
Connect the controller to a switch using a direct
Ethernet connection (LAN). Use a LAN cable
with RJ-45 connectors for this.
Now start the SensorFinder program. You
will find this program on the provided CD.
Click the button
Start Scan
. Select
the designated controller from the list.
In order to change the address settings,
click the button
Change IP...
• Address type: static IP address
• IP address: 169.254.168.150
1
• Subnet mask: 255.255.0.0
Click the button
Apply
to transmit the
changes to the controller.
Click the button
Start Browser
to
connect the controller with your default
browser.
1) Requires that the LAN connection on the
PC uses, for example, the following IP ad-
dress: 169.254.168.1.
Wait until Windows has estab-
lished a network connection
(Connection with limited con-
nectivity).
Now start the Sensor-
Finder program. You will
find this program on the
provided CD.
Click the button
Start
Scan
. Select the desig-
nated controller from the
list.
Click the button
Start
Browser
to connect the
controller with your default
browser.
Enter the controller in the DHCP / register the
controller in your IT department.
The controller gets assigned an IP address from your
DHCP server. You can check this IP address with the
SensorFinder program.
Now start the SensorFinder program.
You will find this program on the provided CD.
Click the button
Start Scan
. Select the des-
ignated controller from the list.
Click the button
Start Browser
to connect
the controller with your default browser.
Alternatively: If DHCP is used and the DHCP server is
linked to the DNS server, access to the controller via
a host name of the structure “KSS6430_<serial_num-
ber>” is possible.
Start a web browser on your PC. To achieve a
controller with the serial number “01234567”,
type in the address bar on your browser
“KSS6430_01234567”.
Interactive web pages for setting the controller and peripherals are now shown in the web browser.