Page 27
Operation
combiSENSOR 64x0
5.2
LED’s on the Controller
DL6430
Range
LED
Color
Function
Range
green
Target within working dis-
tance
red
Working distance exceeded
5.3
Triggering
The measuring value output on combiSENSOR 64x0 is controllable by an external electrical trigger signal or
command. Here, only the digital output is affected.
Triggering release by:
- Trigger input (pin 3 and pin 4 on 4-pole power supply
connector, see Chap.
4.6.2
or
- Software command $GMD, see Chap.
6.4.3
- U
IN
, HIGH
≥
2,0 V
- U
IN
, LOW ≤ 0,8 V
Controller
I = 5 ... 45 mA
100 Ohm
Trigger in
3
4
GND trigger
U trigger
(TTL level,
potential-free)
U =
appr.
1 V
F
Fig. 13 Trigger Input
The trigger type is determined by the command $TRGn, see Chap.
6.4.2
or
- the web interface, see Chap.
6.5.4