Page 101
Operation
colorSENSOR CFO
12
You can choose between Sphere (∆E), Cylinder (∆L; ∆ab), Box (cuboid, ∆L; ∆a; ∆b) and Classifica
-
.
Depending on the distance model selected, it is possible to set the tolerance values (11) to 1 .. 3.
These values describe the size of the tolerance space around the learned color.
Within a color group, the same distance model and the same tolerance values are used for all learned
colors.
No tolerance parameters are set for
Classify
, because the currently measured color is always asso-
ciated to the color saved in the color table with the smallest distance to the measured color
When teaching a new color group, the values selected under Color group settings
(Color
group) are used for the distance model and the tolerance values (11).
Example: Cylinder distance model
Half height (L) = 8
Radius (a/b) = 4
These can be manually adjusted in the edit mode of the group between 0 ... 50.
13
The maximum name length is 64 characters and may consist of only alphanumerical characters (a-z,
A-Z, 0-9, no foreign character attributes), spaces and the special char-#,.().